Briefing

Measuring range: 0.00~14.00pH
Resolution: 0.01pH

Detailed



Quick Details


Place of Origin: Shandong, China (Mainland)Brand Name: Tianyuan OpticalModel Number: ph-01
Measuring range: 0.00~14.00pHResolution: 0.01pHAccuracy: +/-0.05pH
Power Supply: 2*1.5V(Button cell)Operating Temperature: 0℃-80℃Size: 153mm*30mm*15mm
Weight: 52gCalibration: calibration by screwdriverColor: Yellow


Packaging & Delivery


Packaging Details:     Neutral Carton Packaging
Delivery Detail:Shipped in 10 days after payment


SIZE:    100pcs carton,41*19*42cm;       200pcs,41*38*42cm; 

             300pcs carton,57*38*42cm;      400pcs,66*56*40cm




Model: PH-01

Name: High Precision PH Meter




Specification:


*measure range:  0.00pH - 14.00pH

*resolution:    0.01pH

*accuracy:    +/- 0.01pH

*power supply:    2*1.5V (keep using more than 500 hours)

*operating temperature:    0℃ - 80 ℃

*calibration:  two points automatic calibration

*dimension:    153mm*29mm*16mm

*weight:    52g

 

 

Operation:


1. Remove the protective cap

2. First rinse the electrode with distilled water, and suck it with filter paper

3. Turn the meter on by pressing the “ON/OFF” key

4. Immerse the pH meter electrode in the solution to be tested (can not be over the immersion line)

5. Stir gently and wait around 30 seconds till the reading stabilized

6. After finished, clear the electrode with pure water, turn the meter off by pressing the “ON/OFF” key

7. Always replace the protective cap after use

 

 

Calibration:


1. Turn on pH meter

2. Immerse the electrode in pH 6.86 standard buffer solution  (under the temperature of 25 ℃)

3. After digit stabilized, press the “CAL” button for 5 seconds  (the instrument comes into the automatic calibration status)  wait until the display digit is 6.86 pH, then complete pH 6.86 point calibration

4. Clean the electrode with distilled water and dry it filter paper, immerse the electrode in pH 4.00 standard buffer solution

5. After digit stabilized, press “CAL” button for 5 seconds, wait until the digit is 4.00 pH, complete the pH 4.00 point calibration

6. Clean the electrode with distilled water and wipe clean the spare water, put the electrode in the pH 9.18 standard solution until the date stable, the data will be in 9.17pH - 9.19pH, the data is in the range of allowable error

7. This instrument calibration work is completed after all steps

 

 

Attention:


The instrument need re-calibrate in the following condition

Long time no use or use very long time after calibration

The electrode is used very frequently

The testing accuracy requirement is very high

Do not press the “CAL” button long time in the air, otherwise the meter need re-calibrate

 

 

PH electrode maintenance:


If the electrode placed long time,and electrode exposed to air, then showing values is slow and not stable. Because of electrode dry. Just need immerse the electrode into water for few hours.

 


LOW-BATT:


When the display value is fuzzy or not show, it shall promptly replace the battery. Unscrew battery cover and replace the battery. Pay attention to the polarity of battery

 


Warranty:


The instrument maintains for a year, from the delivery date, guaranteeing due to production and material defects. If the instrument need to be changed or repaired (not the artificial damage or caused by fault using), please mail the manual and the instrument together to our factory. We will solve all the problems for you

 

 

Our services

1. we have good after-sales service . we are online  24 hours  .you can contact us any time  if there has problem .

 

2. we can supply you picture modification service . our designer can modify picture for you according to your requirements

 

3.we can print logo on the ph meter/tds&ec meter for you .


Nearest port for product export
Shanghai, Qingdao, Tianjin, Shenzhen
Delivery clauses under the trade mode
FOB
Acceptable payment methods
T/T, PayPal, Westem Union, Escrow
Contact Suppliers

Follow this Suppliers