BT-2001 is the first high performance instrument which can be used under both dry and liquid system.
BT-2001 Laser Particle Size Analyzer Technical Specification
Dry
1. Size range: 0.1-1000μ m
2. Sample adding method: Continual aerosol jet
3. Repeatability: ≤ 3% (CRM D50)
4. Accuracy: ≤ 3% (CRM D50)
5. Principle: Mie Scattering Theory
6. Laser: Imported semiconductor laser
7. Operating system: Windows XP/ Windows 7
8. Interface: RS232 or USB
9. Photocells: 80pcs, forward 72, lateral 8
10. Disperse medium: Compressed air or inert
11. Pressure: 0.1-0.8Mpa, adjustable
12. Air flowrate: 400-6000 L/min
13. Sample required: 5g-20g
BT-2001 Laser Particle Size Analyzer Liquid:
1. Size range: 0.1-1000μ m
2. Repeatability: <1% (CRM D50)
3. Accuracy: <1% (CRM D50)
4. Principle: Mie Scattering Theory
5. Laser: Imported semiconductor laser
6. Ultrasonic Power: 50W
7. Tank Capacity: 500ml
8. Circulating Flowrate: 500-2500ml/min
9. Agitator Speed: 500-2500rpm