Basic Information
|
|
Dimension
|
Height 1720mm/Width 635mm/Thickness 578mm
|
Weight
|
About75kg
|
Electrical Parameters
|
Voltage ~110/220V Frequency 50/60Hz Power 500VA Backup Battery Capacity 2*12V/4AH |
Pressure Monitoring | |
Venous Pressure Monitoring(Vp) | -500mmHg~+700mmHg Accuracy: ±5mmHg |
Arterial Pressure Monitoring(Ap) | -500mmHg~+700mmHg Accuracy: ±5mmHg |
Transmembrane Pressure Monitoring(Tmp) | -500mmHg~+700mmHg Accuracy: ±5mmHg |
Primary Membrane External Pressure Monitoring(Pm1) | -500mmHg~+700mmHg Accuracy: ±5mmHg |
Secondary Membrane Pressure Monitoring(Pm2) | -500mmHg~+700mmHg Accuracy: ±5mmHg |
Liquid Balance Control | |
Scale Number | 4 |
Rehydration Scale | 0kg~30kg Accuracy: ±5g or ±0.1% |
Waste Liquid Scale | 0kg~30kg Accuracy: ±5g or ±0.1% |
Auxiliary Rehydration Scale I | 0kg~12kg Accuracy: ±5g or ±0.1% |
Auxiliary Rehydration Scale I | 0kg~12kg Accuracy: ±5g or ±0.1% |
Flow Monitoring | |
Blood pump flow(BP) | 0mL/min, 30mL/min~600mL/min Accuracy: set value±10% |
LP1 | 0mL/min, 1mL/min~250mL/min Accuracy: ±0.1mL/min or ±5% of set value |
Waste Liquid Pump Flow(UFP) | 0mL/min, 1mL/min~250mL/min Accuracy: ±0.1mL/min or ±5% of set value |
LP2 | 0mL/min, 1mL/min~250mL/min Accuracy: ±0.1mL/min or ±5% of set value |
Function Pump Flow(FP) | 0mL/min, 1mL/min~250mL/min Accuracy: ±0.1mL/min or ±5% of set value |
Infusion pump flow | Flow Adjustable Range: 0mL/h~10mL/h Error Range: ±0.2mL/h Rapid Injection Flow: 10mL/h~2000mL/h Syringe Specification: 10mL/20mL/30mL/50mL |
Dehydration Error | ≤±20mL/h |
Liquid Balance Control | ≤100mL |
Substitution fluid flow temperature control | |
Control Range | 33ºC~40ºC |
Accuracy | Control error±1ºC |
Safety Monitoring | |
Blood Leakage Monitoring | ≤0.35mL/min(HCT32%) |
Air Monitoring | >0.02mL bubbles |
Treat Mode
|
CNME040107
|
||
Citriate anticoagulation
|
No
|
||
Substitution Fluid Double Channel Input
|
No
|
||
CVVH - Continuous Veno-Venous Hemofiltration
|
Yes
|
||
CVVHD - Continuous Veno-Venous Hemodialysis | Yes | ||
CVVHDF - Continuous Veno-Venous Hemodiafiltration | No | ||
SCUF -Slow Continuous Ultrafiltration | Yes | ||
HP - Hemoperfusion | Yes | ||
PE - Plasma Exchange | Yes | ||
PA - Plasma Adsorption | Yes | ||
CPFA - Continuous Plasma Filtration And Adsorption | No | ||
FPSA - Fractional Plasma Separation And Absorption | No | ||
MARS - Molecular Absorbent Recycling System | No | ||
SPAD - Single Pass Albumin Dialysis | No | ||
DFPP - Double Filtration Plasmapheresis | No | ||
RAD - Repeated Pass Albumin Dialysis | No | ||
PDF - Plasmadiafiltration | No |