group nameDefibrillator
Min Order1 piece
modelCNME011308D
payment methodL/C, D/A, D/P, Western Union, MoneyGram, T/T, Paypal
update timeWed, 28 Jul 2021 15:58:09 GMT
Classification Physiological Functions of Diagnosis and Monitorin
Type Defibrillator
Certification CE
Output Biphasic Truncated Exponential
Energy Sequence 150, 200j
Charge Time 8 Sec. to 150j
Min Order1 piece
Model NO.
|
CNME011308D
|
Analysis Time | 9 Sec |
Controls | Two Buttons- on/off, Shock |
Dimension | 303*216*89mm |
Operating Temperature | 0ºC to 40ºC |
Specification | Weight: 2.0kg |
Origin | China |
Output
|
Biphasic Truncated exponential
|
|
Energy Sequence
|
150, 150, 200J
|
|
Charge Time
|
8 sec. to 150J 10sec. to 200J |
|
Analysis Time
|
9 sec.
|
|
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery | Less than 30 seconds | |
The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks. | Less than 35 seconds | |
The maximum time from initially switching power on to readiness for discharge. | Less than 25 seconds | |
Audible Prompts | 20 audible prompts | |
Visual Prompts | LED prompts | |
Controls | Two buttons-On/Off, Shock | |
Output Energy Accuracy | ±15% into any impedance from 25 to 175Ω | |
Output disabled when the patient impedance is outside limits | 20Ω to 200Ω | |
Battery | ||
Part No Non-rechargeable | CR123A-4*2 12V d.c. 3.0Ah | |
Capacity | 100 discharges at 200 Ioules 120 discharges at 150 Joules | |
Shelf Life(25ºC±15ºC) | 10 years(5 years storage + 5 years standby) 5 years standby (after installation) |
网络文化经营许可证:浙ICP备14006100号-11 2007-2024 Tradevv.com. All rights reserved.